Published November 2013 | Version v1
Journal article

The performance determination of a Compton-suppression spectrometer and the measurement of the low level radioactive Samples

Description

The performance of a new Compton-suppression spectrometer consisting of one HPGe detector and three NaI(Tl) detectors was studied. The peak-to-Compton ratio for a 137Cs source is 1150 and the integral background count rate is 0.3 5 s−1 over the energy interval 20–3000 keV. The spectrometer was used to acquire both Compton-suppressed and non-suppressed spectra of aerosol samples collected in Beijing following the Fukushima nuclear accident. - Highlights: • The performance of a Compton-suppression spectrometer (CSS) was determined. • Aerosol samples collected after the Fukushima accident were measured by CSS. • The activity concentrations of fission product were determined in the aerosol samples

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2013.03.045

Additional details

Identifiers

DOI
10.1016/j.apradiso.2013.03.045;
PII
S0969-8043(13)00135-8;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
81
Journal Page Range
p. 128-131
ISSN
0969-8043
CODEN
ARISEF

Conference

Title
6. international conference on radionuclide metrology - Low level radioactivity measurement techniques
Acronym
ICRM-LLRMT'12
Dates
17-21 Sep 2012
Place
Jeju Island (Korea, Republic of)

Optional Information

Copyright
Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.