Validation of accuracy and reliability of a simple measurement device for the lumber range of motion (SMD-L-ROM)
- 1. Dokkyo Medical Univ., School of Medicine, Mibu, Tochigi (Japan)
Description
It is important to evaluate mobility of the lumbar spine for assessment of lumbar spinal disorders and their therapeutic effects. We developed a simple measurement device for the lumbar range of motion (SMD-L-ROM) which can be installed on the trunk. SMD-L-ROM allows to evaluate mobility of the lumbar spine without x-ray exposure. Accuracy and reliability of the measurement device was assessed in this article. SMD-L-ROM comprises two metallic immobilization plates (IPs) with a vertically installed bar and a rubber band for fixation of IP to the trunk. Two IPs were installed at the level of Th12 and the sacrum. Range of motion (ROM) of the lumbar spine was measured by using SMD-L-ROM and radiographic modalities (X-ray and CT). Consistency of the measured values by both of SMD-L-ROM and radiographic modalities was evaluated. Furthermore, inter- and intra-observer agreement of measured values by SMD-L-ROM was assessed. Regarding measurement of ROM, there was a high correlation between SMD-L-ROM and X-ray/CT especially in the sagittal and axial plane. There was no significant difference in the average values of ROM between the two methods. On the other hand, SMD-L-ROM did not detect accurate position of the lumbar spine in all dimensions. There were no inter- and intra-observer errors of measured values by SMD-L-ROM. ROM of the lumbar spine was simply and economically measured with SMD-L-ROM. Accuracy and reliability of SMD-L-ROM was good enough for ROM measurement in the lumbar spine. (author)
Additional details
Publishing Information
- Journal Title
- Dokkyo Journal of Medical Sciences
- Journal Volume
- 34
- Journal Issue
- 2
- Journal Page Range
- p. 85-93
- ISSN
- 0385-5023
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 39037395
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; BIOLOGICAL FUNCTIONS; BIOMEDICAL RADIOGRAPHY; CAT SCANNING; EQUIPMENT; IMAGE PROCESSING; MOTION; RELIABILITY; VALIDATION; VERTEBRAE
- Descriptors DEC
- BODY; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; MEDICINE; NUCLEAR MEDICINE; ORGANS; PROCESSING; RADIOLOGY; SKELETON; TESTING; TOMOGRAPHY