Published January 2005 | Version v1
Journal article

The performance of the INER improved free-air ionization chamber in the comparison of air kerma calibration coefficients for medium-energy X-rays

Description

This paper describes modifications to an original design, correction factors and uncertainty evaluations for an improved free-air ionization chamber constructed at the Institute of Nuclear Energy Research (INER, Taiwan). In addition, a comparison of secondary standard air kerma calibration coefficients for 100-250 kV medium-energy X-rays was performed to verify the experimental accuracy and measurement consistency of the improved chamber. The comparison results showed a satisfactory agreement in the measurements which were within the combined expanded uncertainties (k=2)

Additional details

Identifiers

DOI
10.1016/j.radmeas.2004.03.033;
PII
S1350448704001209;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
39
Journal Issue
1
Journal Page Range
p. 1-10
ISSN
1350-4487
CODEN
RMEAEP

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
36035978
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ACCURACY; AIR; CALIBRATION; CALIBRATION STANDARDS; CORRECTIONS; IONIZATION CHAMBERS; KERMA; PERFORMANCE; TAIWAN
Descriptors DEC
ASIA; CHINA; FLUIDS; GASES; ISLANDS; MEASURING INSTRUMENTS; RADIATION DETECTORS; STANDARDS

Optional Information

Copyright