The performance of the INER improved free-air ionization chamber in the comparison of air kerma calibration coefficients for medium-energy X-rays
Description
This paper describes modifications to an original design, correction factors and uncertainty evaluations for an improved free-air ionization chamber constructed at the Institute of Nuclear Energy Research (INER, Taiwan). In addition, a comparison of secondary standard air kerma calibration coefficients for 100-250 kV medium-energy X-rays was performed to verify the experimental accuracy and measurement consistency of the improved chamber. The comparison results showed a satisfactory agreement in the measurements which were within the combined expanded uncertainties (k=2)
Additional details
Identifiers
- DOI
- 10.1016/j.radmeas.2004.03.033;
- PII
- S1350448704001209;
Publishing Information
- Journal Title
- Radiation Measurements
- Journal Volume
- 39
- Journal Issue
- 1
- Journal Page Range
- p. 1-10
- ISSN
- 1350-4487
- CODEN
- RMEAEP
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 36035978
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; AIR; CALIBRATION; CALIBRATION STANDARDS; CORRECTIONS; IONIZATION CHAMBERS; KERMA; PERFORMANCE; TAIWAN
- Descriptors DEC
- ASIA; CHINA; FLUIDS; GASES; ISLANDS; MEASURING INSTRUMENTS; RADIATION DETECTORS; STANDARDS
Optional Information
- Copyright
- Copyright (c) 2004 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.