Published September 2000 | Version v1
Miscellaneous Open

Characteristics and quality test of X-ray with CZT detector

  • 1. Dept. of Physics and Matematics, Brawijaya University, Malang (Indonesia)

Description

The study examines the use of direct measurements of x-ray spectra for testing the quality of x-ray beam by using a peltier-cooled CZT detector under different conditions. The typical calibration of the spectrometry system shows that the energy resolution of the system is 1,2 keV at 122 keV of 57 Co. The utilization CZT based spectrometer for assessing the quality of x-ray machine on its spectra show that the CZT could accurately measure the spectra at various kVp, m As and filtration, except at the kVp greater than 140 and m As higher than 2 without added filtration. A comparison of CZT with the field instruments showed that there is a reasonable agreement between the Keithley and CZT at lower energies regardless of filtration however at high energies there is a large difference.In contrast the discrepancy between the CZT and Nero increased at lower photon energies particularly for high filtration

Files

34041802.pdf

Files (1.0 MB)

Name Size Download all
md5:b9b4fdf4e596fac640c0292980afb3b4
1.0 MB Preview Download
Part of:
Proceedings of the Eighth Scientific Presentation on Environment and Radiation Safety

Additional details

Additional titles

Original title (Indonesian)
Karakteristik Dan Pengujian Kualitas Sinar-X Dengan Detektor CZT

Publishing Information

Publisher
Centre for Research and Development of Radiation Safety and Nuclear Biomedicine, National Nuclear Energy Agency
Imprint Place
Jakarta (Indonesia)
Imprint Title
Proceedings of the Eighth Scientific Presentation on Environment and Radiation Safety
Imprint Pagination
269 p.
Journal Page Range
p. 199-205
ISSN
0854-4085
Report number
INIS-ID--022

Conference

Title
Environment and Radiation Safety
Original Conference Title
Presentasi Ilmiah Keselamatan Radiasi Dan Lingkungan VIII
Acronym
8. Scientific Presentation
Dates
23-24 Aug 2000
Place
Jakarta (Indonesia)

Optional Information

Notes
12 refs.; tabs.; 5 figs. Imprint:Prosiding Presentasi Ilmiah Keselamatan Radiasi dan Lingkungan VIII