Testing the absolute beam intensity of the high-energy pulsed electron beam with a double-mode readout ionization chamber

  • 1. IHEP of the Chinese Academy of Sciences, Beijing 100049 (China)
  • 2. Shandong University, Shandong 250100 (China)
  • 3. Lanzhou University, Gansu 730000 (China)

Description

We constructed an ionization chamber (IC) to test the absolute intensity of the BEPC-LINAC (Beijing Electron Positron Collider-Linear Accelerator) test beam. The IC was adapted for the 1.89 GeV high-energy electron beam, with pulse time width of 1.2 ns and frequency of 25 Hz, by equipping it with a double-mode readout and choosing the optimum circuit parameters for the readout modes. The measured absolute intensity of the test beam is 7.2x109 electron/s, and is consistent with PSPICE simulations