Beta-gamma discriminator circuit

Description

The major difficulty encountered in the determination of beta-ray dose in field conditions is generally the presence of a relatively high gamma-ray component. Conventional dosimetry instruments use a shield on the detector to estimate the gamma-ray component in comparison with the beta-ray component. More accurate dosimetry information can be obtained from the measured beta spectrum itself. At Los Alamos, a detector and discriminator circuit suitable for use in a portable spectrometer have been developed. This instrument will discriminate between gammas and betas in a mixed field. The portable package includes a 256-channel MCA which can be programmed to give a variety of outputs, including a spectral display, and may be programmed to read dose directly

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02/MF A01; 1 as DE85002378.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
9 p.
Report number
LA-UR--84-3389

Conference

Title
Nuclear science symposium.
Dates
31 Oct - 2 Nov 1984.
Place
Orlando, FL (USA).

Optional Information

Secondary number(s)
CONF-841007--11.