Published August 2012 | Version v1
Journal article

Cone-beam computed tomography images of phantoms with simulating trabecular bone structure fabricated using micro-stereolithography

  • 1. Aichi-Gakuin University, School of Dentistry, Nagoya, Aichi (Japan)
  • 2. Aichi-Gakuin University, Dental Hospital, Nagoya, Aichi (Japan)

Description

Phantoms with simulating mandibular trabecular bone structure fabricated using micro-stereolithography were applied for a quantitative evaluation of cone-beam computed tomography (CT) images. Phantoms with a three-dimensional lattice, in which a square pillar was set at 0.10, 0.20, 0.30 and 0.40 mm at the edge, spaced at 0.60 mm, were made of an acrylic resin using micro-stereolithography. Cone-beam CT with a 0.1 mm-cubic voxel was then performed, and a plot profile of voxel values was analyzed. The plot profiles of seven peaks corresponding to pillars in phantoms with 0.20, 0.30 and 0.40 mm-pillars were visually observed in cone-beam CT images. However, peaks of corresponding pillars in the 0.10 mm-pillar phantom could not be distinguished. The differences between average measured and designed distance were very small. Cone-beam CT images of phantoms with 0.20, 0.30 and 0.40 mm-pillars fabricated using microstereolithography were obtained. (author)

Availability note (English)

Available from http://dx.doi.org/10.2535/ofaj.89.27

Additional details

Identifiers

Publishing Information

Journal Title
Okajimas Folia Anatomica Japonica (Online)
Journal Volume
89
Journal Issue
2
Journal Page Range
p. 27-33
ISSN
1881-1736

Optional Information

Notes
23 refs., 6 figs., 2 tabs.