Description
The paper describes the principle of measuring digitally the flow velocity of local voids, and discusses the accuracy of this method. It is shown that the spectrum of local void velocities can be determined reliably with digital values. The measurement errors of void velocity ug (m/sec) are within about ±9% in cases when the probes have detector head lengths of 0.8~1.0mm, and head diameters of 0.1 mmφ, when the voids have a flow velocity in the range of 0.2 m/sec≤ug ≤1.2 m/sec and a radius of curvature of void of 2mm≤R≤16mm.
Additional details
Identifiers
- DOI
- 10.3327/jnst.9.613;
Publishing Information
- Journal Title
- Journal of Nuclear Science and Technology
- Journal Volume
- 9
- Journal Issue
- 10
- Series
- J. Nucl. Sci. Technol. (Tokyo).
- Journal Page Range
- 613-617
- ISSN
- 1881-1248
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 4059405
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; DATA; DIGITAL CIRCUITS; ELECTRIC PROBES; ERRORS; GATING CIRCUITS; MEASURING METHODS; RELIABILITY; SIGNALS; TWO-PHASE FLOW; VELOCITY; VOIDS
- Descriptors DEC
- ELECTRONIC CIRCUITS; FLUID FLOW; INFORMATION; PROBES