Evaluating computer-aided detection algorithms
- 1. Department of Radiology, University of Pittsburgh, Pittsburgh, Pennsylvania 15261 (United States)
- 2. Department of Radiology, University of Michigan, Ann Arbor, Michigan 48109 (United States)
Description
Computer-aided detection (CAD) has been attracting extensive research interest during the last two decades. It is recognized that the full potential of CAD can only be realized by improving the performance and robustness of CAD algorithms and this requires good evaluation methodology that would permit CAD designers to optimize their algorithms. Free-response receiver operating characteristic (FROC) curves are widely used to assess CAD performance, however, evaluation rarely proceeds beyond determination of lesion localization fraction (sensitivity) at an arbitrarily selected value of nonlesion localizations (false marks) per image. This work describes a FROC curve fitting procedure that uses a recent model of visual search that serves as a framework for the free-response task. A maximum likelihood procedure for estimating the parameters of the model from free-response data and fitting CAD generated FROC curves was implemented. Procedures were implemented to estimate two figures of merit and associated statistics such as 95% confidence intervals and goodness of fit. One of the figures of merit does not require the arbitrary specification of an operating point at which to evaluate CAD performance. For comparison a related method termed initial detection and candidate analysis was also implemented that is applicable when all suspicious regions are reported. The two methods were tested on seven mammography CAD data sets and both yielded good to excellent fits. The search model approach has the advantage that it can potentially be applied to radiologist generated free-response data where not all suspicious regions are reported, only the ones that are deemed sufficiently suspicious to warrant clinical follow-up. This work represents the first practical application of the search model to an important evaluation problem in diagnostic radiology. Software based on this work is expected to benefit CAD developers working in diverse areas of medical imaging
Additional details
Identifiers
- DOI
- 10.1118/1.2736289;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 34
- Journal Issue
- 6
- Journal Page Range
- p. 2024-2038
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38101719
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ALGORITHMS; BIOMEDICAL RADIOGRAPHY; COMPUTER CODES; DETECTION; IMAGES; LICENSES; MAMMARY GLANDS; OPTIMIZATION; SENSITIVITY ANALYSIS
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; GLANDS; MATHEMATICAL LOGIC; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY
Optional Information
- Notes
- (c) 2007 American Association of Physicists in Medicine