Published August 2017 | Version v1
Miscellaneous Open

Differences in radiation dosage value (CTDIw) and image noise on sequence and spiral engineering in computer tomography face bone (study on modality of scan brand Siemens 6 slice)

  • 1. Poltekkes Kemenkes, Semarang (Indonesia)
  • 2. Mahasiswa Prodi D-IV Teknik Radiologi (Indonesia)

Description

This Research use parameter of exposure factor are tube current and time (mAs) of CT scan for measuring dose radiation (CTDIw) and image noise with sequence and spiral techniques. The purpose of this research is to know difference of radiation dose (CTDIw) and image noise with sequence and spiral techniques of variation mAs at CT Scan Siemens 6 slice. This research was an experimental study. A five various use of mAs (from 250 mAs up to 290 mAs) and then effect of radiation doses (CTDIw) and image noise due to the techniques being implemented were compared and measured with tested to use paired T-Test to take a conclusion with trust storey level 95 % (ρ value <0,05). That having meaning that there is difference of dose value at sequence and spiral techniques, while test of Wilcoxon at show image noise assess significantly with ρ value equal to 0,000 (ρ value <0,05). It means there are difference of noise image between of sequence and spiral techniques. (author)

Files

51070161.pdf

Files (342.1 kB)

Name Size Download all
md5:8501b375f70a27c52e64ba99a335026c
342.1 kB Preview Download
Part of:
Proceedings of the seminar on nuclear safety 2017: Development of nuclear power oversight based on public information and communication technology

Additional details

Additional titles

Original title (Indonesian)
Perbedaan indikator nilai dosis radiasi (CTDIw) dan image noise pada teknik sekuens dan spiral pada computed tomography face bone (studi pada modalitas CT scan merk siemens 6 slice)

Publishing Information

Publisher
Nuclear Energy Regulatory Agency
Imprint Place
Yogyakarta (Indonesia)
Imprint Title
Proceedings of the seminar on nuclear safety 2017: Development of nuclear power oversight based on public information and communication technology
Imprint Pagination
389 p.
Journal Page Range
p. 190-194
ISSN
1412-3258
Report number
INIS-ID--89

Conference

Title
The Seminar on Nuclear Safety 2017
Original Conference Title
Seminar Keselamatan Nuklir 2017
Dates
1 Aug 2017
Place
Yogyakarta (Indonesia)

INIS

Country of Publication
Indonesia
Country of Input or Organization
Indonesia
INIS RN
51070161
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CAT SCANNING; COMPUTERIZED TOMOGRAPHY; IMAGES; NOISE; RADIATION DOSES; RADIATION PROTECTION; STRUCTURAL CHEMICAL ANALYSIS
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DOSES; TOMOGRAPHY

Optional Information

Notes
26 refs.; 7 tabs.; 3 figs. Imprint:Prosiding seminar keselamatan nuklir 2017: Pengembangan pengawasan ketenaganukliran berbasis teknologi informasi dan komunikasi publik