Published October 2006 | Version v1
Report Open

Safety Analysis Report for Ir-192 HDR Sealed Source

Description

Safety evaluation to transport safely Ir-192 HDR source for RALS, which is produced at HANARO research reactor in KAERI, was carried out. The specimen of the sealed source was deformed as a little scratched in impact tests. The diameter of the sealed source was reduced as 0.01 mm in percussion test. However, the sealed source after each test would not be shattered. In heat test, the sealed source was not deformed and its surface was changed from a silver color to a brass yellow, but was not melted or dispersed. In leakage tests, which was conducted before and after each test, the leakage in the sealed sources did not happen. Therefore the Ir-192 HDR sources are designed and fabricated in accordance with the safety criteria of the special form and the sealed sources specified on the related regulations and codes

Availability note (English)

Available from INIS in electronic form; Also available from Korea Atomic Energy Research Institute, Taejon (Korea, Republic of)

Files

38056816.pdf

Files (1.4 MB)

Name Size Download all
md5:fdac58a4e774a31888bbf2203b7d90e8
1.4 MB Preview Download

Additional details

Publishing Information

Imprint Pagination
46 p.
Report number
KAERI/TR--3142/2006

Optional Information

Notes
7 refs, 20 figs, 5 tabs