Published August 2006 | Version v1
Journal article

Effect of concomitant analytes on As signal during pre-evaporation of the solvent prior to introduction into inductively coupled plasma mass spectrometry

  • 1. Department of Chemistry, Queen's University, Kingston, ON., K7L 3N6 (Canada)

Description

Axial profiling was done, with detection by inductively coupled plasma time-of-flight mass spectrometry, to investigate how a pre-evaporation extension interface (i.e. heated glass tube between the nebulizer and the torch) affects the signal from As and other analytes. A reproducible axial profile was observed for As+ in a 20-μg/L multielemental solution in 1% HNO3, with a signal enhancement ratio (heated/unheated) of 1.55 (obtained by taking the ratio of the signal observed at the optimal axial position in each case). In contrast, the axial profile of a 20-μg/L monoelemental As solution also in 1% HNO3 showed a signal intensity ratio (heated/unheated) of 0.65 (very similar to that for Se). No such difference in signal was observed for all the other analytes investigated, which are ionic in solution, in contrast to As and Se, which would be mostly in neutral form in 1% HNO3. In fact, for ionic analytes, the optimal signal upon heating was similar or higher than that without heating, with an inverse dependency on analyte mass. Increasing the number of concomitant analytes and/or their concentration enhanced the As+ signal upon heating the interface. Switching to a 0.01 M NaOH matrix, which would not evaporate and would dominate the ionic strength even if As is then entirely in anionic form, similarly resulted in enhancement of the As signal upon heating the extension. All these observations could be rationalized by changes in the number of Coulomb fission events occurring during pre-evaporation, which depend on the size and charge of droplets, as well as the identity and concentration of the matrix

Additional details

Identifiers

DOI
10.1016/j.sab.2006.08.012;
PII
S0584-8547(06)00253-9;

Publishing Information

Journal Title
Spectrochimica Acta. Part B, Atomic Spectroscopy
Journal Volume
61
Journal Issue
8
Journal Page Range
p. 965-970
ISSN
0584-8547
CODEN
SAASBH

Optional Information

Copyright