Published May 1, 2009 | Version v1
Journal article

3-D dosimetric evaluation of 2.5 mm HD120 multileaf system for intensity modulated stereotactic radiosurgery using optical CT based polymer gel dosimetry

  • 1. Department of Radiation Oncology, Columbia University, New York, NY (United States)

Description

A Trilogy TX equipped with a 2.5 mm HD120 multileaf collimator system is available for the treatment of radiosurgery and IMRT. In this study, we evaluated the 3-D dosimetric impact of leaf width on an IMRT radiosurgery plan by comparing the target coverage and the dose gradient around the target, produced from both a 2.5 mm HD120 high-definition MLC system and a 5mm-leaf-width millennium 120 MLC system, using an optical CT based polymer gel dosimetry system. The 2.5 mm MLC improves target conformity and surrounding tissue sparing when compared to that of 5 mm MLC.

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/164/1/012056

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
164
Journal Issue
1
Journal Page Range
[6 p.]
ISSN
1742-6596

Conference

Title
5. international conference on radiotherapy gel dosimetry
Dates
29 Sep - 3 Oct 2008
Place
Hersonissos, Crete (Greece)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41058027
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
DOSIMETRY; GELS; POLYMERS; RADIATION DOSES; RADIOTHERAPY; SURGERY
Descriptors DEC
COLLOIDS; DISPERSIONS; DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY