The accuracy of frameless stereotactic intracranial radiosurgery
Creators
- 1. Department of Radiation Oncology, VU University Medical Center, Amsterdam (Netherlands)
Description
Purpose: To determine the accuracy of frameless stereotactic radiosurgery using the BrainLAB ExacTrac system and robotic couch by measuring the individual contributions such as the accuracy of the imaging and couch correction system, the linkage between this system and the linac isocenter and the possible intrafraction motion of the patient in the frameless mask. Materials and methods: An Alderson head phantom with hidden marker was randomly positioned 31 times. Automated 6D couch shifts were performed according to ExacTrac and the deviation with respect to the linac isocenter was measured using the hidden marker. ExacTrac-based set-up was performed for 46 patients undergoing hypofractionated stereotactic radiotherapy for 135 fractions, followed by verification X-rays. Forty-three of these patients received post-treatment X-ray verification for 79 fractions to determine the intrafraction motion. Results: The hidden target test revealed a systematic error of 1.5 mm in one direction, which was corrected after replacement of the system calibration phantom. The accuracy of the ExacTrac positioning is approximately 0.3 mm in each direction, 1 standard deviation. The intrafraction motion was 0.35 ± 0.21 mm, maximum 1.15 mm. Conclusion: Intrafraction motion in the BrainLAB frameless mask is very small. Users are strongly advised to perform an independent verification of the ExacTrac isocenter in order to avoid systematic deviations.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.radonc.2010.06.012Additional details
Identifiers
- DOI
- 10.1016/j.radonc.2010.06.012;
- PII
- S0167-8140(10)00399-3;
Publishing Information
- Journal Title
- Radiotherapy and Oncology
- Journal Volume
- 97
- Journal Issue
- 3
- Journal Page Range
- p. 390-394
- ISSN
- 0167-8140
- CODEN
- RAONDT
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 42081099
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACCURACY; ERRORS; HEAD; MOTION; PHANTOMS; POSITIONING; RADIOTHERAPY; SURGERY
- Descriptors DEC
- BODY; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; STRUCTURAL MODELS; THERAPY
Optional Information
- Copyright
- Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.