The accuracy of frameless stereotactic intracranial radiosurgery

  • 1. Department of Radiation Oncology, VU University Medical Center, Amsterdam (Netherlands)

Description

Purpose: To determine the accuracy of frameless stereotactic radiosurgery using the BrainLAB ExacTrac system and robotic couch by measuring the individual contributions such as the accuracy of the imaging and couch correction system, the linkage between this system and the linac isocenter and the possible intrafraction motion of the patient in the frameless mask. Materials and methods: An Alderson head phantom with hidden marker was randomly positioned 31 times. Automated 6D couch shifts were performed according to ExacTrac and the deviation with respect to the linac isocenter was measured using the hidden marker. ExacTrac-based set-up was performed for 46 patients undergoing hypofractionated stereotactic radiotherapy for 135 fractions, followed by verification X-rays. Forty-three of these patients received post-treatment X-ray verification for 79 fractions to determine the intrafraction motion. Results: The hidden target test revealed a systematic error of 1.5 mm in one direction, which was corrected after replacement of the system calibration phantom. The accuracy of the ExacTrac positioning is approximately 0.3 mm in each direction, 1 standard deviation. The intrafraction motion was 0.35 ± 0.21 mm, maximum 1.15 mm. Conclusion: Intrafraction motion in the BrainLAB frameless mask is very small. Users are strongly advised to perform an independent verification of the ExacTrac isocenter in order to avoid systematic deviations.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radonc.2010.06.012

Additional details

Identifiers

DOI
10.1016/j.radonc.2010.06.012;
PII
S0167-8140(10)00399-3;

Publishing Information

Journal Title
Radiotherapy and Oncology
Journal Volume
97
Journal Issue
3
Journal Page Range
p. 390-394
ISSN
0167-8140
CODEN
RAONDT

INIS

Country of Publication
Ireland
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
42081099
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; ERRORS; HEAD; MOTION; PHANTOMS; POSITIONING; RADIOTHERAPY; SURGERY
Descriptors DEC
BODY; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; STRUCTURAL MODELS; THERAPY

Optional Information

Copyright
Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.