Published November 1, 2010 | Version v1
Journal article

2D AND 3D dose verification at The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital using EPIDs

  • 1. Department of Radiation Oncology, The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital, Plesmanlaan 121, 1066 CX Amsterdam (Netherlands)

Description

A review is given of the clinical use of EPID dosimetry in the Department of Radiation Oncology of The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital. All curative plans (almost all IMRT or VMAT) are verified with EPID dosimetry, mostly in vivo. The 2D approach for IMRT verification and the 3D method for VMAT verification are elucidated and their clinical implementation described. It has been shown that EPID dosimetry plays an important role in the total chain of verification procedures that are implemented in our department. It provides a safety net for advanced treatments such as IMRT and VMAT, as well as a full account of the dose delivered.

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/250/1/012020

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
250
Journal Issue
1
Journal Page Range
[6 p.]
ISSN
1742-6596

Conference

Title
6. international conference on 3D radiation dosimetry
Acronym
IC3DDose
Dates
22-26 Aug 2010
Place
Hilton Head Island, SC (United States)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
42053539
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
DOSIMETRY; HOSPITALS; IN VIVO; NEOPLASMS; RADIATION DOSES; SAFETY; VERIFICATION
Descriptors DEC
BUILDINGS; DISEASES; DOSES; MEDICAL ESTABLISHMENTS