Self-shielding corrections in cylindrical samples in gamma spectrometry with germanium well-type detectors

  • 1. Instituto Balseiro, CNEA and Universidad Nacional de Cuyo, Bustillo 9500, 8400 Bariloche (Argentina)
  • 2. UAIN, GAEN, Centro Atómico Bariloche, Comisión Nacional de Energía Atómica (CNEA), Bustillo 9500, 8400 Bariloche (Argentina)

Description

Self-shielding in spatially distributed samples can be very significant when detecting photons below 100 keV. A correction method has been developed for cylindrical samples, typically used in measurements of natural radioactivity with well-type detectors, as is the example of sediment dating. The method calculates the probability of photons to escape the sample, using Monte Carlo techniques with a program written in C. The effects of self-shielding on the angular distribution of photons have been indirectly analyzed as a function of sample geometry, given a fixed detector geometry. The results given by the program have also been used to provide a straightforward way of calculating the self-shielding factors, with uncertainties. Required inputs are the attenuation coefficient of the sample, its radius, and its height. A procedure has also been specified in order to make good use of all the information. To check on the method, measurements of reference material of known activity have been compared with calculated values, obtaining very satisfactory results

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2012.12.085

Additional details

Identifiers

DOI
10.1016/j.nima.2012.12.085;
PII
S0168-9002(12)01633-6;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
705
Journal Page Range
p. 132-139
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.