Description

Using the Monte Carlo code MCNP-5, the photons and photoneutrons spectra generated in the head stock of the lineal accelerator (Linac) Varian of 15 MV of the Cancerology State of Nayarit were determined. For the calculations a heterogeneous head stock was modeled, more compatible with the work conditions. In the center of the head stock a tungsten target was located on a copper support, followed by the flattened filter. The photons and photoneutrons spectra were obtained accelerating electrons and making them collide against the target to produce photons by Bremsstrahlung, these photons were transported inside the head stock and the photons and photoneutrons spectra were calculated in a punctual detector located under the flattened filter and in the isocenter. The spectra were evaluated in punctual detectors that were located in the plane from the isocenter to the long of the X and Y axes each 20 cm, in an equidistant way, up to 2 m, so much in the longitudinal and transversal axes. In the calculations were used histories 5E(6) with the purpose of obtaining smaller uncertainties to 1%. It was found that the photons spectrum in the punctual detector inside the head stock presents a pick of 1.25 MeV in the energy interval of 0.5 and 1.5 MeV, later suffers a filtration and diminishes in asymptote form. This spectrum modifies when the beam reaches the isocenter, diminishing the low energy photons. Inside the head stock the photoneutrons spectrum shows a structure with two picks, one before 1 MeV and other after 1 MeV; this is for effect of the collimators geometry and the distance. Finally an increment of the total neutrons flow to 60 cm of distance of the isocenter on the Y axis was observed, due to the design geometry of the modeling heterogeneous head stock. (Author)

Files

43009434.pdf

Files (670.8 kB)

Name Size Download all
md5:52327b098c4c8936a074029b1e3e6dc8
670.8 kB Preview Download

Additional details

Additional titles

Original title (Spanish)
Espectros de fotones y fotoneutrones de un LINAC de 15 MV

Publishing Information

Publisher
Sociedad Mexicana de Irradiacion y Dosimetria
Imprint Place
Mexico D. F. (Mexico)
Imprint Pagination
[10 p.]
Report number
INIS-MX--2631

Conference

Title
12. International Symposium; 22. National Congress on Solid State Dosimetry
Original Conference Title
12. Conferencia Internacional. 22. Congreso Nacional sobre Dosimetria de Estado Solido
Dates
5-9 Sep 2011
Place
Mexico D. F. (Mexico)