Depth of interaction in scintillation crystal by light collimation for the design of open-quotes universal nuclear medicine imagerclose quotes
Creators
- 1. Univ. Clinical Center and Vinca Institute for Nuclear Sciences, Belgrade (Yugoslavia)
Description
The detector head for a dual headed open-quotes Universal Nuclear Medicine Imagerclose quotes (UNMI) is designed. It is based on position sensitive (PS) area detectors fitted with a thick (0.75 inches) NaI(Tl) crystal. In order to make it open-quotes universalclose quotes, a means for determining the depth of interaction (DOI) in the crystal is devised. Crystal thickness is divided into two layers, each 3/8 inches thick, by insertion of a thin (0.03 inches) quartz (glass) layer; these layers added to the external lower and upper surfaces of crystal, too, form two internal open-quotes light collimatorclose quotes (LC) channels by their reflectively (specular) polished couplings. These DOI channels, according to the light refraction / reflection law direct the light reflected from quartz layers (as exceeding a critical angle) through NaI(Tl) layers to the external parts of LC system; these are connected via quartz optical cables to the DOI determining PM tubes. DOI electronics including a coincidence circuit, connecting signals from conventional PS and DOI PM tubes arrays, identifies a DOI layer and shifts an, x, y, address to the particular DOI memory stack. The UNMI design is likely to retain spatial resolution of PS detectors - gamma cameras for low energy single photon emitters and to improve spatial resolution in PET studies done by PS detectors fitted with thick NaI crystals, enabling a proper correction for parallax error
Additional details
Publishing Information
- Publisher
- IEEE Service Center.
- Imprint Place
- Piscataway, NJ (United States)
- Imprint Title
- 1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3
- Imprint Pagination
- 2138 p.
- Journal Page Range
- p. 1175-1180.
Conference
- Title
- Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
- Dates
- 2-9 Nov 1996.
- Place
- Anaheim, CA (United States).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 28067990
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- COLLIMATORS; DESIGN; GAMMA CAMERAS; IMAGE PROCESSING; NUCLEAR MEDICINE; POSITRON COMPUTED TOMOGRAPHY; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY; SODIUM IODIDES; SOLID SCINTILLATION DETECTORS
- Descriptors DEC
- ALKALI METAL COMPOUNDS; CAMERAS; COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; HALIDES; HALOGEN COMPOUNDS; IODIDES; IODINE COMPOUNDS; MEASURING INSTRUMENTS; MEDICINE; RADIATION DETECTORS; SCINTILLATION COUNTERS; SODIUM COMPOUNDS; TOMOGRAPHY