Published 2019 | Version v1
Report

A Study on the Monte Carlo Calculation of Kilovoltage X-ray Beam Backscatter Factors

  • 1. Institute of Medical Physics, School of Physics, University of Sydney NSW 2006 (Australia)
  • 2. Chris O'Brien Lifehouse, 119 – 143 Missenden Road, Camperdown NSW 2050 (Australia)
  • 3. Department of Medical Radiation Physics and Nuclear Medicine, Karolinska University Hospital, SE-171 76 Stockholm (Sweden)

Description

Backscatter factors in water (Bw) are important dosimetric quantities used in reference and relative dosimetry of kilovoltage (kV) radiotherapy x-ray beams. Bw is defined as the ratio of water kerma on the beam central axis at a point at the surface of a water phantom to the water kerma at the same point free in air. It is not straightforward to measure Bw and data are typically determined from Monte Carlo (MC) calculations; these are found in dosimetry protocols for kilovoltage x-ray beams [5]. Published values may, however, be for x-ray beams quite different from the beams found in the clinic or laboratory. In addition, the publication of ICRU Report 90 providing updated physics data, which may impact the values of Bw, has been one rationale for a new database of Bw values calculated by Andreo for an extensive number of x-ray beam spectra, field sizes and source-to-surface distances. In this work, independent calculations of Bw for a broad range of x-ray beam qualities and field sizes using Monte Carlo techniques has been made for clinical beams generated by Pantak SXT150 and DXT300 generators. The results have been compared with those from the new database of Andreo, based on the Penelope MC system.

Part of:
International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry (IDOS 2019). Book of Extended Synopses

Additional details

Publishing Information

Imprint Title
International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry (IDOS 2019). Book of Extended Synopses
Imprint Pagination
455 p.
Journal Page Range
p. 168-169
Report number
IAEA-CN--273

Conference

Title
International Symposium on Standards, Applications and Quality Assurance in Medical Radiation Dosimetry
Acronym
IDOS 2019
Dates
18-21 Jun 2019
Place
Vienna (Austria)

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
53119578
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BEAMS; DISTANCE; DOSIMETRY; ICRU; KERMA; MONTE CARLO METHOD; PHANTOMS; RADIOTHERAPY; X RADIATION
Descriptors DEC
CALCULATION METHODS; ELECTROMAGNETIC RADIATION; INTERNATIONAL ORGANIZATIONS; IONIZING RADIATIONS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIATIONS; RADIOLOGY; STRUCTURAL MODELS; THERAPY

Optional Information

Notes
8 refs., 1 fig.
Secondary number(s)