Published August 2006 | Version v1
Miscellaneous

Measurement system of femtosecond pulse radiolysis based on photocathode linac

  • 1. Osaka Univ., Inst. of Scientific and Industrial Research, Ibaraki, Osaka (Japan)

Description

Femto second pulse radiolysis system is developed in ISIR using a laser photo-cathode RF gun S-band LINAC. For achievement of femto second time resolution, we developed the measurement system called 'Equivalent velocity spectroscopy'. Degradation of time resolution by velocity difference between an electron and light in a sample was avoided by the equivalent velocity spectroscopy. In the equivalent velocity spectroscopy, analyzing light incident on a sample with angle θ(cos θ=1/n), and the wave front of an electron pulse was adjusted. (author)

Part of:
The 3rd annual meeting of Particle Accelerator Society of Japan and the 31th Linear Accelerator Meeting in Japan. Proceedings

Additional details

Identifiers

Publishing Information

Imprint Title
The 3rd annual meeting of Particle Accelerator Society of Japan and the 31th Linear Accelerator Meeting in Japan. Proceedings
Imprint Pagination
819 p.
Journal Page Range
p. 562-564

Conference

Title
3. annual meeting of Particle Accelerator Society of Japan; 31. Linear Accelerator Meeting in Japan
Dates
2-4 Aug 2006
Place
Sendai, Miyagi (Japan)

Optional Information

Notes
1 ref., 9 figs.