Published June 1, 1984 | Version v1
Journal article

Determination of neutron fluence based on the measurement of isotopic variation of cadmium and gadolinium

  • 1. Comissao Nacional de Energia Nuclear de Brasil, Sao Paulo. Inst. de Pesquisas Energeticas e Nucleares

Description

The determination of neutron fluence using mass spectrometric measurements of the isotopic composition variation of elements with high neutron capture cross section has been attempted. Cadmium and gadolinium have been used as monitors. The results indicate that the fluence values obtained using the variation in the isotopic ratio 114Cd/113Cd present a standard deviation of the order of 15%, whereas values calculated using 158Gd/157Gd show a deviation of about 5%. The fluence values calculated by the two methods agree within about 10%. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. A
Journal Volume
223
Journal Issue
1
Series
CODEN: NIMRD.;Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
137-140
ISSN