A method for determination of parameters of the initial electron beam hitting the target in linac

  • 1. Department of Radiation Oncology, Dong Nai General Hospital, Bien Hoa City, Dong Nai Province (Viet Nam)
  • 2. Chi Anh Med. Tech. Co.,Ltd., Ho Chi Minh City (Viet Nam)
  • 3. Faculty of Physics and Engineering Physics, VNUHCM-University of Science, Ho Chi Minh City (Viet Nam)

Description

Detailed characteristics of the electron beam incident on the target of the Linac are almost impossible to be measured. With Monte Carlo technique it is possible to simulate the transport of radiation through the accelerator head and find out the initial beam parameters. These parameters, such as energy spectrum and spatial distribution can be ascertained by matching the simulated dose distributions with the measured dose distributions using trial and error. In this work, a Monte Carlo modeling of the HPD Siemens Primus linear accelerator in 6 MV photon beams at Dong Nai general hospital was performed. The simulated Percent depth doses (PDD) and beam profiles (OCR) were then compared with the measured ones. Excellent agreements were obtained between simulations and measurements with an average difference of 0.7% for PDD less than 2% for OCR. The percentage gamma passing rate is 100% with 1% dose difference and 1 mm distance to agreement as acceptance criteria. The best suited energy and radius of the electron beam incident on the target were found at 6.04 MeV and at FWHM = 1.2 mm respectively. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/851/1/012032

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
851
Journal Issue
1
Journal Page Range
[6 p.]
ISSN
1742-6596

Conference

Title
10. international seminar on medical physics; 1. AMDI international oncology symposium
Dates
26-28 Aug 2016
Place
Penang (Malaysia)