Published January 15, 1978 | Version v1
Journal article

An air light guide, plastic scintillation counter

  • 1. LAPP, Annecy-le-Vieux, France

Description

A sheet (30x30cm2) of plastic scintillator is viewed by a PM XP 2020, without any solid light guide. The accuracy achieved in measuring time of flight of particles is 520 ps. The simplicity of such a counter makes possible the construction of big hodoscope systems with good resolution times. (Auth.)

Additional details

Identifiers

Publishing Information

Journal Title
Nuclear Instruments and Methods
Journal Volume
148
Journal Issue
2
Series
Nucl. Instrum. Methods.
Journal Page Range
359-362
ISSN
0029-554X

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent