Published June 1997 | Version v1
Journal article

Intravascular stent graft with polyurethane and metallic stent: experimental study

  • 1. Samsaung Medical Center, Seoul (Korea, Republic of)
  • 2. Seoul National Univ. College of Medicine, Seoul (Korea, Republic of)

Description

To evaluate the usefulness of a new model of the stent graft, and of tissue response related to placement of the stent graft. The stent graft was constructed from polyurethane (Pellethane) graft and Hanaro stent(12mm in diameter, 45mm in length, 10 bends). A stent grafts was inserted into the lower thoracic aorta in each of six adult mongrel dogs(body weight, 12-16kg). At one, two, four, and six months, follow-up studies of angiography and spiral CT angiography were preformed to evaluate wascular patency, vascular stenosis, and thrombus formation. Two dogs were sacrificed at 1month, 2months, and 6months after insertion of the stent graft and macroscopic, light microscopic, and scanning electron microscopic examinations of the aortic segment including the stent graft were performed to evaluate intimal hyperplasia, endothelial growth to the graft, and thrombus formation. During follow-up at one, two, four, and six months, angiography or spiral CT angiography showed 20-100% luminal stenosis or occlusion of the lower thoracic aorta by the thrombus and perigraft leaks in three dogs(50%), and collateral vessels caused by occlusion of the aorta in two (33.3%). On gross examination, there were thrombi of 1-5mm thickness at the graft portions in all dogs, and this thickness gradually increased. The mean thickness of intimal hyperplasia at the stent portion gradually increased from 120μm to 227μm and the mean thickness of intimal hyperplasia at the graft portion from 93μm to 914μm. This thickness was greater at the graft portion than at the stent portion. Scanning electron microscopy showed elliptical endothelial lining on the neointimal surfaces at each end of the graft. Thrombi caused stenosis or occlusion of the stent graft. In order for such a graft to be ideal, further study is needed

Additional details

Publishing Information

Journal Title
Journal of the Korean Radiological Society
Journal Volume
36
Journal Issue
6
Series
40 refs., 7 figs., 3 tabs
Journal Page Range
p. 955-964
ISSN
0301-2867