The design of a focused ultrasound transducer array for the treatment of stroke: a simulation study

  • 1. Department of Imaging Research, Sunnybrook Research Institute, Toronto (Canada)

Description

High intensity focused ultrasound (HIFU) is capable of mechanically disintegrating blood clots at high pressures. Safe thrombolysis may require frequencies higher than those currently utilized by transcranial HIFU. Since the attenuation and focal distortion of ultrasound in bone increases at higher frequencies, resulting focal pressures are diminished. This study investigated the feasibility of using transcranial HIFU for the non-invasive treatment of ischemic stroke. The use of large aperture, 1.1–1.5 MHz phased arrays in targeting four clinically relevant vessel locations was simulated. Resulting focal sizes decreased with frequency, producing a maximum –3 dB depth of field and lateral width of 2.0 and 1.2 mm, respectively. Mean focal gains above an order of magnitude were observed in three of four targets and transducer intensities required to achieve thrombolysis were determined. Required transducer element counts are about an order of magnitude higher than what currently exists and so, although technically feasible, new arrays would need to be developed to realize this as a treatment modality for stroke. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/57/15/4951

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
57
Journal Issue
15
Journal Page Range
p. 4951-4968
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
Cuba
INIS RN
43128257
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ATTENUATION; BLOOD; CONTAINERS; SIMULATION; SKELETON; TRANSDUCERS
Descriptors DEC
BIOLOGICAL MATERIALS; BODY; BODY FLUIDS; MATERIALS; ORGANS