Detection efficiency of electron multipliers for low-energy electrons

Description

Pulse-height spectra have been obtained for electron multipliers with discrete dynodes of CuBe. A peak in the spectrum shows up at low pulse heights (≈ 0.1 pC for a multiplier with a dc gain of about 106), but a long tail extends out to very large pulses. These observations can be understood in terms of secondary emission processes. They have provided a good basis for the design and calibration of a soft particle spectrometer for the ISIS-2 satellite. The detection efficiency for 300-eV electrons, with a 14-stage venetian blind multiplier having a total gain of 106, is about 80%. Below 100 eV the detection efficiency falls off, but in practice this can be overcome by postacceleration. At higher energies it also decreases to about 30% at 4 keV. This efficiency is reduced at very high counting rates, above 107 sec−1, owing partly to amplifier bandwidth limitations related to noise immunity requirements, and partly to the finite rise time of the multiplier pulses.

Additional details

Identifiers

Publishing Information

Journal Title
Journal of Applied Physics
Journal Volume
43
Journal Issue
4
Series
J. Appl. Phys.
Journal Page Range
1585-1590
ISSN
0021-8979

INIS

Optional Information

Notes