Detection efficiency of electron multipliers for low-energy electrons
Creators
Description
Pulse-height spectra have been obtained for electron multipliers with discrete dynodes of CuBe. A peak in the spectrum shows up at low pulse heights (≈ 0.1 pC for a multiplier with a dc gain of about 106), but a long tail extends out to very large pulses. These observations can be understood in terms of secondary emission processes. They have provided a good basis for the design and calibration of a soft particle spectrometer for the ISIS-2 satellite. The detection efficiency for 300-eV electrons, with a 14-stage venetian blind multiplier having a total gain of 106, is about 80%. Below 100 eV the detection efficiency falls off, but in practice this can be overcome by postacceleration. At higher energies it also decreases to about 30% at 4 keV. This efficiency is reduced at very high counting rates, above 107 sec−1, owing partly to amplifier bandwidth limitations related to noise immunity requirements, and partly to the finite rise time of the multiplier pulses.
Additional details
Identifiers
- DOI
- 10.1063/1.1661366;
Publishing Information
- Journal Title
- Journal of Applied Physics
- Journal Volume
- 43
- Journal Issue
- 4
- Series
- J. Appl. Phys.
- Journal Page Range
- 1585-1590
- ISSN
- 0021-8979
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 3029461
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BETA SPECTROMETERS; DOSE RATES; EFFICIENCY; ELECTRON MULTIPLIERS; ELECTRONS; ENERGY; EV RANGE; KEV RANGE; PULSES; SPECTRA
- Descriptors DEC
- ELECTRON TUBES; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; LEPTONS; MEASURING INSTRUMENTS; SPECTROMETERS
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent