Published February 1, 2013 | Version v1
Journal article

A new external microbeam system in Fudan University

  • 1. Applied Ion Beam Physics Laboratory, Institute of Modern Physics, Fudan University, Shanghai 200433 (China)

Description

A cost-effective and removable external beam system is set up based on the Oxford Microbeam system in Fudan University. In our external beam system, 7.5-μm-thick Kapton film is used as exit window with a diameter of 3.5 mm. The spatial resolution is about 18 μm full width at half maximum (FWHM) on a copper grid. As an example, calcium distribution in otolith is present by the external micro-PIXE. In addition, little change can be done to the external system mentioned above for radiobiology experiments. The exit window can be changed from the focal plane to the observation window of vacuum chamber. By calculation, the beam spot size can reach less than 30 μm. Since the Oxford type octagonal target chamber is popular among the nuclear microprobe facilities, this method can be provided to easily replace the in-vacuum system with the external system, extending the in-vacuum analysis to external beam analysis

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nimb.2012.10.034

Additional details

Identifiers

DOI
10.1016/j.nimb.2012.10.034;
PII
S0168-583X(12)00671-4;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section B, Beam Interactions with Materials and Atoms
Journal Volume
296
Journal Page Range
p. 82-86
ISSN
0168-583X
CODEN
NIMBEU

Optional Information

Copyright