Published May 1977 | Version v1
Report

Drift chamber

  • 1. Kyoto Univ. (Japan). Faculty of Science

Description

Drift chamber is becoming an important detector in high energy physics as a precision and fast position detector because of its high spatial resolution and count-rate. The basic principle is that it utilizes the drift at constant speed of electrons ionized along the tracks of charged particles towards the anode wire in the nearly uniform electric field. The method of measuring drift time includes the analog and digital ones. This report describes about the construction of and the application of electric field to the drift chamber, mathematical analysis on the electric field and equipotential curve, derivation of spatial resolution and the factor for its determination, and selection of gas to be used. The performance test of the chamber was carried out using a small test chamber, the collimated β source of Sr-90, and 500 MeV/C electron beam from the 1.3 GeV electron synchrotron in the Institute of Nuclear Study, University of Tokyo. Most chambers to date adopted one dimensional read-out, but it is very advantageous if the two dimensional read-out is feasible with one chamber when the resolution in that direction is low. The typical methods of delay line and charge division for two dimensional read-out are described. The development of digital read-out system is underway, which can process the signal of a large scale drift chamber at high speed. (Wakatsuki, Y.)

Additional details

Publishing Information

Imprint Title
Proceedings of the symposium on wire chambers and data acquisition with CAMAC standard
Journal Page Range
p. 128-147.
Report number
KEK--77-2

Conference

Title
Symposium on wire chambers and data acquisition with CAMAC standard.
Dates
2 - 3 Sep 1976.
Place
Oho, Ibaraki, Japan.