Determination of carbon in stainless steel and plutonium using infrared detection of carbon dioxide

Creators

  • 1. Los Alamos National Lab., NM (USA)

Description

Carbon in stainless steel and plutonium metal in the 3-1000 ppm range is burned at 13000C in a stream of purified oxygen and the carbon dioxide produced is quantified by integrating the signal from an infrared detector. The relative standard deviation is 10% or better. This method is approximately five times faster than the conventional manometric method and is equally accurate and precise. Use of a platinum resistance furnace makes the method more suitable for glove boxes or shielded enclosures than methods requiring inductively heated furnaces. (Auth.)

Additional details

Publishing Information

Journal Title
Anal. Chim. Acta
Journal Volume
130
Journal Issue
1
Series
Anal. Chim. Acta.
Journal Page Range
215-220