Published 1978 | Version v1
Book

Gamma-ray dosimetry errors with thermoluminescent dose meters

  • 1. California Univ., Livermore (USA). Lawrence Livermore Lab.

Description

Gamma-ray dosimetry errors can occur when thermoluminescence dose meters (TLDs) are irradiated in thin-walled containers. An over-response has been observed for LiF TLDs, irradiated by an isotropic source, of approximately 25% for 24Na; 20% for 60Co, and 2% for 137Cs. The over-response for CaF2 TLDs was found to be less than the over-response of LiF TLDs. Under-response errors can occur for TLDs in thin-walled containers irradiated under narrow-beam geometry conditions because electron equilibrium is not established, but this error is well known and is not discussed in this paper. The observed over-response is attributable to electrons originating in the radiation source holder and the surrounding air that are scattered into the TLD material. This over-response is less pronounced for 137Cs gamma rays because few scattered electrons have sufficient energy to penetrate the TLD; however, for higher energy gamma rays, more electrons can panetrate the TLD, resulting in a greater dose. This effect will also be observed with any dose meter that has a small sensitive volume surrounded by a thin wall. The over-response of a dosimetry film in its paper wrapping was measured to be approximately 25% for 60Co gamma rays and 3% for 137Cs gamma rays, although this would rarely introduce an error since film dose meters are usually irradiated in a thick-walled badge. This paper presents the measured over-responses of various TLD chips for several exposure conditions and experimental results which demonstrate the cause of the over-response. (author)

Additional details

Publishing Information

Publisher
IAEA.
Imprint Place
Vienna
ISBN
92-0-010578-5
Imprint Title
National and international standardization of radiation dosimetry
Series
Proceedings series.
Journal Page Range
v. 2 p. 335-347.

Conference

Title
International symposium on national and international standardization of radiation dosimetry.
Dates
5 - 9 Dec 1977.
Place
Atlanta, GA, USA.

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
11523149
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference, Numerical Data
Descriptors DEI
ERRORS; EXPERIMENTAL DATA; GAMMA DOSIMETRY; GRAPHS; ISOLATED VALUES; THERMOLUMINESCENT DOSEMETERS
Descriptors DEC
DATA; DATA FORMS; DOSEMETERS; DOSIMETRY; INFORMATION; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; NUMERICAL DATA

Optional Information

Secondary number(s)
IAEA-SM--222/15.