A method to calculate the transmission function (luminosity) of mini-orange magnetic spectrometer systems is described. Inhomogeneities of the toroidal magnetic field, fringe-field effects, as well as luminosity losses due to backscattering effects at the detector surface are found to have a significant influence on the transmission of the system and are included in the calculations. Very good agreement between measured and calculated transmission functions is obtained. (orig.)
Additional details
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research, Section A
- Journal Volume
- 275
- Journal Issue
- 2
- Series
- Nucl. Instrum. Methods Phys. Res., Sect. A.
- Journal Page Range
- 301-306
- ISSN
- 0168-9002
- CODEN
- NIMAE
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 20030435
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- BACKSCATTERING; BEAM LUMINOSITY; CHARGED PARTICLE DETECTION; ENERGY DEPENDENCE; FLAT MAGNETIC SPECTROMETERS; INHOMOGENEOUS FIELDS; KEV RANGE 100-1000; MAGNETIC FIELDS; MEV RANGE 01-10; POSITION SENSITIVE DETECTORS; TRAJECTORIES; TRANSMISSION
- Descriptors DEC
- DETECTION; ENERGY RANGE; KEV RANGE; MAGNETIC SPECTROMETERS; MEASURING INSTRUMENTS; MEV RANGE; RADIATION DETECTION; RADIATION DETECTORS; SCATTERING; SPECTROMETERS