Published March 1979 | Version v1
Journal article

Measurement of retained xenon in advanced fuels by microprobe analysis

Creators

  • 1. Commission of the European Communities, Karlsruhe (Germany, F.R.). European Inst. for Transuranium Elements

Description

The author presents a procedure for direct quantitative analysis of bonded xenon in advanced fuel by EPMA. The quality of the data obtained with the method is illustrated by a radial xenon concentration profile from a section of mixed carbonitride fuel. The main features of this profile and the principal sources of error in the measured concentrations are briefly described. (Auth.)

Additional details

Publishing Information

Journal Title
J. Nucl. Mater.
Journal Volume
80
Journal Issue
1
Series
J. Nucl. Mater.
Journal Page Range
190-193
ISSN