Published 2011 | Version v1
Journal article

Correction of the volumetric effect in a 2D-array of ionisation chambers

Description

There are three main snags related to the measurement of spatial distributions of absorbed dose with two-dimensional array devices of ionisation chambers: spatial under-sampling, volumetric effect and inaccuracy in electric charge to absorbed dose conversion coefficient due to lack of lateral electronic equilibrium. In this work we have experimentally estimated the error in the raw measurement of dose by one of such chambers. We have found that the major error factor is the volumetric effect, which can be corrected through the well-known techniques of convolution/deconvolution. (Author) 14 refs.

Availability note (English)

Available www.sefm.es

Additional details

Additional titles

Original title (Spanish)
Correccion del efecto volumetrico en una matriz 2D de camaras de ionizacion

Publishing Information

Journal Title
Revista de Fisica Medica
Journal Volume
12
Journal Issue
1
Journal Page Range
p. 33-38
ISSN
1576-6632

INIS

Country of Publication
Spain
Country of Input or Organization
Spain
INIS RN
42068278
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Descriptors DEI
DOSE RATES; EQUILIBRIUM; IONIZATION; RADIATION DOSES; RADIATION EFFECTS; RADIATION PROTECTION; VOLUMETRIC ANALYSIS
Descriptors DEC
CHEMICAL ANALYSIS; DOSES; QUANTITATIVE CHEMICAL ANALYSIS