Published 2011
| Version v1
Journal article
Correction of the volumetric effect in a 2D-array of ionisation chambers
Description
There are three main snags related to the measurement of spatial distributions of absorbed dose with two-dimensional array devices of ionisation chambers: spatial under-sampling, volumetric effect and inaccuracy in electric charge to absorbed dose conversion coefficient due to lack of lateral electronic equilibrium. In this work we have experimentally estimated the error in the raw measurement of dose by one of such chambers. We have found that the major error factor is the volumetric effect, which can be corrected through the well-known techniques of convolution/deconvolution. (Author) 14 refs.