Published July 1982
| Version v1
Journal article
Improvement of image contrast in x-ray transmission micrographs of organism with a titanium tube
- 1. Teikyo Univ., Tokyo (Japan)
Description
To improve the image contrast in X-ray transmission micrographs of organisms, a fine-focus sealed-off tube with a titanium target has been built, which can produce quasi monochromatic X-rays at the K sub(α) characteristic wavelength of 2.7 A. The intensity ratio of the characteristic ray to continuous ones reaches a maximum when a voltage of 8 kV is applied to the tube. Micrographs of organisms obtained with this tube show up a better image contrast than those taken with a copper-anode tube. (author)
Additional details
Publishing Information
- Journal Title
- Oyo Butsuri
- Journal Volume
- 51
- Journal Issue
- 7
- Series
- Oyo Butsuri.
- Journal Page Range
- 848-853
- ISSN
- 0369-8009
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 15025740
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANODES; BERYLLIUM; BIOMEDICAL RADIOGRAPHY; COPPER; FABRICATION; IMAGES; K ABSORPTION; KEV RANGE 01-10; SOFT X RADIATION; TITANIUM; X-RAY RADIOGRAPHY; X-RAY TUBES
- Descriptors DEC
- ABSORPTION; ALKALINE EARTH METALS; DIAGNOSTIC TECHNIQUES; ELECTRODES; ELECTROMAGNETIC RADIATION; ELECTRON TUBES; ELEMENTS; ENERGY RANGE; EQUIPMENT; IONIZING RADIATIONS; KEV RANGE; MEDICINE; METALS; RADIATIONS; TRANSITION ELEMENTS; X RADIATION; X-RAY EQUIPMENT