Published 1985
| Version v1
Book
Direct contact applicator with convergent effect of EM field for microwave hyperthermia
- 1. Dept. of Electrical Engineering, Faculty of Science and Technology, Keio Univ., 3-14-1 Hiyoshi, Kouhoku-ku, Yokohama, 223
Description
A new type applicator for microwave hyperthermia operated at 433 MHz is proposed. This applicator, which is a waveguide array type, has a convergent effect of the electromagnetic field radiated from the applicator, and focal length of it can be changeable. The electric field distributions in the model of human tissues from the aperture of the applicator were measured, and the simulation of temperature distributions was performed in the model heated with the applicator
Additional details
Publishing Information
- Publisher
- IEEE Service Center.
- Imprint Place
- Piscataway, NJ (USA)
- Imprint Title
- Digest of the proceedings of the 1985 IEEE-MTTS international microwave symposium
- Journal Page Range
- p. 86-89.
Conference
- Title
- IEEE MTT-S international microwave symposium.
- Dates
- 4-6 Jun 1985.
- Place
- St. Louis, MO (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 18027709
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ELECTROMAGNETIC FIELDS; HEATING; HYPERTHERMIA; MAN; MHZ RANGE 100-1000; MICROWAVE EQUIPMENT; MICROWAVE RADIATION; OPERATION; PERFORMANCE TESTING; RADIATION DETECTION; SIMULATION; SPECIFICATIONS; TEMPERATURE DISTRIBUTION; TISSUES; WAVEGUIDES
- Descriptors DEC
- ANIMALS; BODY; BODY TEMPERATURE; DETECTION; ELECTROMAGNETIC RADIATION; ELECTRONIC EQUIPMENT; EQUIPMENT; FREQUENCY RANGE; MAMMALS; MHZ RANGE; PRIMATES; RADIATIONS; TESTING; VERTEBRATES