Effects of interfering constituents on tritium smears

Description

Tritium smears are performed by Health Protection Operations (HPO) to assess transferable contamination on work place surfaces, materials for movement outside Radiologically Controlled Areas (RCA), and product containers being shipped between facilities. Historically, gas proportional counters were used to detect transferable tritium contamination collected by smearing. Because tritium is a low-energy beta emitter, gas proportional counters do not provide the sensitivity or the counting efficiency to accurately measure the tritium activity on the smear. Liquid Scintillation Counters (LSC) provide greater counting efficiency for the low-energy beta particles along with greater reliability and reproducibility compared to gas flow proportional counters. The purpose of this technical evaluation was to determine the effects of interfering constituents such as filters, dirt and oil on the counting efficiency and tritium recoveries of tritium smears by LSC

Availability note (English)

MF available from INIS under the Report Number; Also available from OSTI as DE94014225; NTIS; US Govt. Printing Office Dep.

Files

25074534.pdf

Files (548.5 kB)

Name Size Download all
md5:cde1df2d3f2f557b8bf3f6f37132f505
548.5 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
7 p.
Report number
WSRC-TR--93-475

Optional Information

Contract/Grant/Project number
Contract AC09-89SR18035
Funding organization
USDOE, Washington, DC (United States).