Published May 7, 2014 | Version v1
Journal article

Analysis of the electron temperature measurement in TCABR tokamak by electron cyclotron emission and infrared Thomson scattering diagnostics

  • 1. Institute of Physics, University of São Paulo, SP (Brazil)
  • 2. Associaçào EURATOM/IST, Instituto de Plasmas e Fusão Nuclear – Laboratório Associado, Instituto Superior Técnico, 1049-001 Lisboa (Portugal)

Description

This work presents the experimental analysis of the central electron temperature measured by the electron cyclotron emission (ECE) radiometer and the infrared Thomson Scattering (ITS) diagnostic. The detection of the ECE radiation is done by a heterodyne scanning radiometer that works at the second harmonic extraordinary mode, in frequency range from 50 to 85GHz, which allows measurement of the radial profile of electron temperature with good spatial and temporal resolutions. The ITS diagnostic uses a Neodymium Glass laser (wavelength 1.054 μm). This ITS diagnostic measures the electron temperature in the center of plasma column one time during plasma shot. Results also show a discrepancy between the two diagnostics in the electron temperature measurement in the presence of Magnetohydrodynamics activity that gives an explanation for this apparent inconsistency.

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/511/1/012039

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
511
Journal Issue
1
Journal Page Range
[6 p.]
ISSN
1742-6596

Conference

Title
15. international congress on plasma physics; LAWPP2010: 13. Latin American workshop on plasma physics
Acronym
ICPP2010
Dates
8-13 Aug 2010
Place
Santiago (Chile)