Optimization of pencil beam widths for use in an electron pencil beam algorithm
- 1. and Department of Medical Physics, Cancer Treatment and Research Foundation of Nova Scotia, Halifax, Nova Scotia B3H 1V7
- 2. Department of Clinical Physics, Princess Margaret Hospital, Toronto, Ontario Canada, M4X 1K9 and Department of Medical Biophysics, University of Toronto, Toronto, Ontario, Canada M4X 1K9
Description
Pencil beam algorithms for the calculation of electron beam dose distributions have come into widespread use. These algorithms, however, have generally exhibited difficulties in reproducing dose distributions for small field dimensions or, more specifically, for those conditions in which lateral scatter equilibrium does not exist. The work described here has determined that this difficulty can arise from the manner in which the width of the pencil beam is calculated. A unique approach for determining the pencil beam widths required to accurately reproduce small field dose distributions in a homogeneous phantom is described and compared with measurements and the results of other calculations. This method has also been extended to calculate electron beam dose distributions in heterogeneous media and the results of this work are presented. Suggestions for further improvements are discussed
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 15
- Journal Issue
- 4
- Series
- Med. Phys.
- Journal Page Range
- 489-497
- ISSN
- 0094-2405
- CODEN
- MPHYA
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 19088555
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ALGORITHMS; BEAM PROFILES; ELECTRON DOSIMETRY; MONTE CARLO METHOD; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIOTHERAPY; WIDTH
- Descriptors DEC
- DIMENSIONS; DOSIMETRY; MEDICINE; MOCKUP; STRUCTURAL MODELS; THERAPY