Description

Background: Seed source has become a popular treatment option in the management of various tumors, particularly in the prostate. Purpose: The aim is to develop accurate and reliable dosimetric parameters that could be used to measure the dose delivered to organs at risk. Methods: Dosimetric parameters (dose rate constant, radial dose function and anisotropy function) of model 6711 125I seed source were calculated with MCNP5 and EGSnrc MC codes following AAPM TG43U1 recommendations. Results: The two results were compared with the relative data recommend by AAPM TG43U1, and the data were as follows: dose rate constant with MCNP5 was in agreement with 0.62%, while that with EGSnrc was 2.07%; radial dose function with MCNP5 was within 0.15%-5.12%, while that with EGSnrc was within 0%-2.48%. Conclusion: The results of two MC codes are in accordance with the recommendations. But that with EGSnrc MC code is better. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Techniques
Journal Volume
37
Journal Issue
2
Journal Page Range
[4 p.]
ISSN
0253-3219

Optional Information

Notes
2 figs., 2 tabs., 6 refs. 020201-1-020201-4