Published 2005 | Version v1
Computer medium

Computational voxel phantom, associated to anthropometric and anthropomorphic real phantom for dosimetry in human male pelvis radiotherapy

  • 1. Minas Gerais Univ., Belo Horizonte, MG (Brazil). Programa de Pos-graduacao em Ciencias e Tecnicas Nucleares

Description

This paper addresses a computational model of voxels through MCNP5 Code and the experimental development of an anthropometric and anthropomorphic phantom for dosimetry in human male pelvis brachytherapy focusing prostatic tumors. For elaboration of the computational model of the human male pelvis, anatomical section images from the Visible Man Project were applied. Such selected and digital images were associated to a numeric representation, one for each section. Such computational representation of the anatomical sections was transformed into a bi-dimensional mesh of equivalent tissue. The group of bidimensional meshes was concatenated forming the three-dimensional model of voxels to be used by the MCNP5 code. In association to the anatomical information, data from the density and chemical composition of the basic elements, representatives of the organs and involved tissues, were setup in a material database for the MCNP-5. The model will be applied for dosimetric evaluations in situations of irradiation of the human masculine pelvis. Such 3D model of voxel is associated to the code of transport of particles MCNP5, allowing future simulations. It was also developed the construction of human masculine pelvis phantom, based on anthropometric and anthropomorphic dates and in the use of representative equivalent tissues of the skin, fatty, muscular and glandular tissue, as well as the bony structure.This part of work was developed in stages, being built the bony cast first, later the muscular structures and internal organs. They were then jointly mounted and inserted in the skin cast. The representative component of the fatty tissue was incorporate and accomplished the final retouchings in the skin. The final result represents the development of two important essential tools for elaboration of computational and experimental dosimetry. Thus, it is possible its use in calibrations of pre-existent protocols in radiotherapy, as well as for tests of new protocols, besides comparative studies among different clinical protocols. (author)

Part of:
Study of semiconductor detectors applied to diagnostic X-ray

Additional details

Additional titles

Original title (Portuguese)
Fantoma computacional de voxel, associado a fantoma real antropomorfico antropometrico, para dosimetria em radioterapia de pelve masculina

Publishing Information

ISBN
85-99141-01-5
Imprint Title
Proceedings of the INAC 2005: International nuclear atlantic conference. Nuclear energy reducing global warming; 14. Brazilian national meeting on reactor physics and thermal hydraulics; 7. Brazilian national meeting on nuclear applications
Imprint Pagination
[4886 p.]
Journal Page Range
[6 p.]

Conference

Title
International nuclear atlantic conference. Nuclear energy reducing global warming; 14. Brazilian national meeting on reactor physics and thermal hydraulics; 7. Brazilian national meeting on nuclear applications
Acronym
INAC 2005
Dates
28 Aug - 2 Sep 2005
Place
Santos, SP (Brazil)

INIS

Country of Publication
Brazil
Country of Input or Organization
Brazil
INIS RN
36114094
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
COMPUTERIZED SIMULATION; M CODES; MALES; MONTE CARLO METHOD; PELVIS; PHANTOMS; RADIOTHERAPY; TISSUE-EQUIVALENT MATERIALS
Descriptors DEC
BODY; CALCULATION METHODS; COMPUTER CODES; MATERIALS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; RADIOLOGY; SIMULATION; STRUCTURAL MODELS; THERAPY

Optional Information

Notes
11 refs., 7 figs. Code: 2256.pdf Imprint:Published only in CD-Rom