Published June 1990
| Version v1
Report
Restricted
Improvement of the bandwidth of the transient digitizers in the LIDAR Thomson scattering diagnostic on JET
Description
The main limitation on the spatial resolution of the LIDAR Thomson scattering diagnostic on the JET tokamak is due to the narrow bandwidth of the detection system. The transient digitizers, Tektronik 7912AD, are the main contributors to the narrow bandwidth. It is shown how the digitizers can be modified to improve the response time from approx. 480 to 410 ps. (author)
Availability note (English)
MF available from INIS under the Report Number; Also available from Risoe Library, DK-4000 Roskilde.
Files
Additional details
Publishing Information
- ISBN
- 87-550-1648-0
- Imprint Pagination
- 22 p.
- Report number
- RISO-M--2873
INIS
- Country of Publication
- Denmark
- Country of Input or Organization
- Denmark
- INIS RN
- 22010848
- Subject category
- S70: PLASMA PHYSICS AND FUSION TECHNOLOGY;
- Descriptors DEI
- DIAGNOSTIC TECHNIQUES; DIGITIZERS; JET TOKAMAK; THOMSON SCATTERING; TRANSIENTS
- Descriptors DEC
- CLOSED PLASMA DEVICES; INELASTIC SCATTERING; SCATTERING; THERMONUCLEAR DEVICES; TOKAMAK DEVICES