Noiseless direct detection of electrons in Medipix2 for electron microscopy
Creators
- 1. MRC Laboratory of Molecular Biology, Hills Road, Cambridge CB2 2QH (United Kingdom)
- 2. CERN, ETT/TT, 1211 Geneva 23 (Switzerland)
Description
We describe the application of a hybrid pixel detector (HPD), Medipix2, containing 256x256 pixels to electron microscopy. Our previous tests on HPDs were made with Medipix1, which was an earlier version of the present detector. Apart from an increase in the size and number of pixels, the main advantage of Medipix2 for our applications is in the ability to set a fairly high threshold voltage. We show that Medipix2 has a high detection efficiency for 120 keV electrons, with no readout noise. The spatial resolution is also excellent as measured by comparing the power spectrum obtained by imaging a standard electron microscopy grid: at Nyquist frequency the power is 76% compared to film. The special properties of Medipix2 allow electron detection in single pixels. No radiation damage was detectable up to a dose of ∼6 Mrads-equivalent to many years normal usage in the electron microscope
Additional details
Identifiers
- DOI
- 10.1016/j.nima.2005.03.022;
- PII
- S0168-9002(05)00615-7;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 546
- Journal Issue
- 1-2
- Journal Page Range
- p. 160-163
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 6. international workshop on radiation imaging detectors
- Dates
- 25-29 Jul 2004
- Place
- Glasgow (United Kingdom)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37040035
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- EFFICIENCY; ELECTRON DETECTION; ELECTRON MICROSCOPES; ELECTRON MICROSCOPY; ELECTRONS; GRIDS; KEV RANGE 100-1000; RADIATION DETECTORS; SPATIAL RESOLUTION; USES
- Descriptors DEC
- CHARGED PARTICLE DETECTION; DETECTION; ELECTRODES; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; KEV RANGE; LEPTONS; MEASURING INSTRUMENTS; MICROSCOPES; MICROSCOPY; RADIATION DETECTION; RESOLUTION
Optional Information
- Copyright
- Copyright (c) 2005 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.