Published March 2022 | Version v1
Journal article

Experimental Evaluation of Scattered X-Ray Spectra due to X-Ray Therapeutic and Diagnosis Equipment for Eye Lens Dosimetry of Medical Staff

  • 1. Japan Atomic Energy Agency, Ibaraki, Japan (Japan)
  • 2. University of Occupational and Environmental Health, Fukuoka (Japan)

Description

For proper monitoring of the eye lens dose, an appropriate calibration factor of a dosimeter and information about the mean energies of X-rays are indispensable. The scattered X-ray energy spectra should be well characterized in medical practices where eye lenses of medical staffs might be high. Scattered X-ray energy spectra were experimentally derived for three different types of X-ray diagnostic and therapeutic equipment, i.e., the computed tomography (CT) scan, the angiography and the fluoroscopy. A commercially available CdZnTe (CZT) spectrometer with a lead collimator was employed for the measurement of scattered X-rays, which was performed in the usual manner. From the obtained energy spectra, the mean energies of the scattered X-rays lied between 40 and 60 keV. This also agreed with that obtained by the conventional half value layer method. The scattered X-rays to which medical workers may be exposed in the region around the eyes were characterized by means of spectrometry. The obtained mean energies of the scattered X-rays were found to match the flat region of the dosimeter response

Additional details

Publishing Information

Journal Title
Journal of Radiation Protection and Research (Online)
Journal Volume
47
Journal Issue
1
Journal Page Range
p. 39-49
ISSN
2466-2461

Optional Information

Notes
21 refs, 6 figs, 6 tabs