Published 2011 | Version v1
Book

Jaws calibration method to get a homogeneous distribution of dose in the junction of hemi fields

Description

Hemi fields treatments are widely used in radiotherapy. Because the tolerance established for the positioning of each jaw is 1 mm, may be cases of overlap or separation of up to 2 mm. This implies heterogeneity of doses up to 40% in the joint area. This paper presents an accurate method of calibration of the jaws so as to obtain homogeneous dose distributions when using this type of treatment. (Author)

Additional details

Additional titles

Original title (Spanish)
Metodo de calibracion de mandibulas para conseguir una distribucion homogenea de dosis en la zona de union de hemicampos