Published April 2002 | Version v1
Report Open

The effect of x-ray summing in calibration of extended energy range GE detectors

Creators

  • 1. STUK - Radiation and Nuclear Safety Authority, Helsinki (Finland)

Description

Gamma spectrometric analyses of low-level samples are usually performed using close measuring geometry. In this measuring set-up the effect of the true coincidence summing (TCS) has to be taken into account. The detection of true coincidence sum pulse is possible when a nuclide emits two or more cascading photons within the resolving time of a spectrometer. The magnitude of summing depends strongly on the total and full energy peak efficiencies of the detector and on the decay scheme of the nuclide. In addition to summing of gamma rays, other radiation, especially X-rays, can be in true coincidence with the gamma rays. X-rays in cascade with gamma rays can originate from electron capture, positron decay, or internal conversion. The summing effects with most of the X-rays can be ignored for the p-type Ge detectors having thick dead layer. On the contrary, the detectors with a thin entrance window, e.g. n-type detectors and new extended range p-type detectors, can experience severe summing of X-rays and gamma rays. (au)

Files

33071277.pdf

Files (60.6 kB)

Name Size Download all
md5:1bb3e1371cc8aae60724cb8bf3d97e19
60.6 kB Preview Download
Part of:
Proceedings of the 8. Nordic seminar on radioecology

Additional details

Publishing Information

ISBN
87-7893-126-6
Imprint Title
Proceedings of the 8. Nordic seminar on radioecology
Imprint Pagination
233 p.
Journal Page Range
p. 206-209
Report number
NKS--70

Conference

Title
8. Nordic seminar on radioecology
Dates
25-28 Feb 2001
Place
Rovaniemi (Finland)

INIS

Country of Publication
Denmark
Country of Input or Organization
Denmark
INIS RN
33071277
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CALIBRATION; COINCIDENCE SPECTROMETRY; GE SEMICONDUCTOR DETECTORS; X RADIATION
Descriptors DEC
COINCIDENCE METHODS; COUNTING TECHNIQUES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; SEMICONDUCTOR DETECTORS

Optional Information

Contract/Grant/Project number
Contract NKS/BOK-2