SU-F-BRE-06: Evaluation of Patient CT Dose Reconstruction From 3D Diode Array Measurements Using Anthropomorphic Phantoms

  • 1. The University of Alabama at Birmingham, Birmingham, AL (United States)
  • 2. The University of Texas MD Anderson Cancer Center, Houston, TX (United States)

Description

Purpose: To compare 3D reconstructed dose of IMRT plans from 3D diode array measurements with measurements in anthropomorphic phantoms. Methods: Six IMRT plans were created for the IROC Houston (RPC) head and neck (H and N) and lung phantoms following IROC Houston planning protocols. The plans included flattened and unflattened beam energies ranging from 6 MV to 15 MV and both static and dynamic MLC tecH and Niques. Each plan was delivered three times to the respective anthropomorphic phantom, each of which contained thermoluminescent dosimeters (TLDs) and radiochromic films (RCFs). The plans were also delivered to a Delta4 diode array (Scandidos, Uppsala, Sweden). Irradiations were done using a TrueBeam STx (Varian Medical Systems, Palo Alto, CA). The dose in the patient was calculated by the Delta4 software, which used the diode measurements to estimate incident energy fluence and a kernel-based pencil beam algorithm to calculate dose. The 3D dose results were compared with the TLD and RCF measurements. Results: In the lung, the average difference between TLDs and Delta4 calculations was 5% (range 2%–7%). For the H and N, the average differences were 2.4% (range 0%–4.5%) and 1.1% (range 0%–2%) for the high- and low-dose targets, respectively, and 12% (range 10%-13%) for the organ-at-risk simulating the spinal cord. For the RCF and criteria of 7%/4mm, 5%/3mm, and 3%/3mm, the average gamma-index pass rates were 95.4%, 85.7%, and 76.1%, respectively for the H and N and 76.2%, 57.8%, and 49.5% for the lung. The pass-rate in the lung decreased with increasing beam energy, as expected for a pencil beam algorithm. Conclusion: The H and N phantom dose reconstruction met the IROC Houston acceptance criteria for clinical trials; however, the lung phantom dose did not, most likely due to the inaccuracy of the pencil beam algorithm in the presence of low-density inhomogeneities. Work supported by PHS grant CA10953 and CA81647 (NCI, DHHS)

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
41
Journal Issue
6
Journal Page Range
p. 392
ISSN
0094-2405
CODEN
MPHYA6

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
46108059
Subject category
S60: APPLIED LIFE SCIENCES;
Descriptors DEI
ALGORITHMS; CAT SCANNING; LUNGS; PHANTOMS; RADIATION DOSES; THERMOLUMINESCENT DOSEMETERS
Descriptors DEC
BODY; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DOSEMETERS; DOSES; LUMINESCENT DOSEMETERS; MATHEMATICAL LOGIC; MEASURING INSTRUMENTS; MOCKUP; ORGANS; RESPIRATORY SYSTEM; STRUCTURAL MODELS; TOMOGRAPHY

Optional Information

Notes
(c) 2014 American Association of Physicists in Medicine