A method for measuring the nuclear transformation rate of 109Cd

  • 1. Radiochemical Centre Ltd., Amersham (UK)

Description

109Cd + sup(109m)Ag sources are measured in a large 4π proportional counter containing argon-methane at atmospheric pressure. The dimensions of the counter were chosen so that the internal conversion electrons from the 88-keV transition are totally absorbed in the gas. The internal conversion electron pulses are counted above a discrimination level selected to exclude K X-ray and Auger electron pulses from electron capture decay. Good resolution in the pulse height spectrum was observed for sources made at the centre of large diameter gold-coated VYNS foils. The calculated nuclear transformation rate is not critically dependent on the value used for the total internal conversion coefficient. (author)

Additional details

Identifiers

Publishing Information

Journal Title
The International Journal of Applied Radiation and Isotopes
Journal Volume
28
Journal Issue
9
Series
Int. J. Appl. Radiat. Isot.
Journal Page Range
773-780