Published April 1, 1999
| Version v1
Journal article
An NaI(Tl) spectrometer system for keV neutron radiative-capture reactions
Creators
- 1. Research Laboratory for Nuclear Reactors, Tokyo Institute of Technology, O-okayama, Meguro, Tokyo 152-8550 (Japan)
- 2. Department of Applied Physics, Tokyo Institute of Technology, O-okayama, Meguro, Tokyo 152-8551 (Japan)
Description
An NaI(Tl) spectrometer system has been installed to measure the cross section of a radiative neutron-capture reaction of a nucleus at an astrophysically relevant energy of between 10 and 500 keV. The system consists of two large anti-Compton NaI(Tl) spectrometers and a new data-taking system. The spectrometer can detect a discrete γ-ray emitted promptly from a neutron-capture state to its low-lying state, and the data-taking system can transfer events with much higher rates, about 30-times higher, compared to the existing system
Additional details
Identifiers
- PII
- S0168900298013643;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 425
- Journal Issue
- 1-2
- Journal Page Range
- p. 302-319
- ISSN
- 0168-9002
- CODEN
- NIMAER
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Syrian Arab Republic
- INIS RN
- 34064004
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CAPTURE; CROSS SECTIONS; GAMMA RADIATION; GAMMA SPECTROMETERS; KEV RANGE 10-100; KEV RANGE 100-1000; NAI DETECTORS
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; ENERGY RANGE; IONIZING RADIATIONS; KEV RANGE; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; SPECTROMETERS
Optional Information
- Copyright