Published January 2017 | Version v1
Journal article

Absolute calibration of a three-axis SQUID-cascade vector magnetometer

  • 1. Leibniz Institute of Photonic Technology, 07745 Jena (Germany)
  • 2. Supracon AG, An der Lehmgrube 11, 07751 Jena (Germany)
  • 3. Department für Geo- und Umweltwissenschaften, Ludwig-Maximilians Universität, Theresienstrasse 41, 80333 München (Germany)

Description

We report on the absolute calibration of a three-axis SQUID-cascade vector magnetometer suited for mobile operation in the Earth's magnetic field. It is based on low temperature superconductor (LTS) dc SQUIDs with sub-micrometer sized cross-type Josephson junctions and exhibits a white noise level of about 10 fT Hz−1/2. Due to the cascaded operation principle, the huge dynamic range required to resolve a measurement range of more than 50 µ T with sub-pT resolution is split onto several electronic channels of less demanding precision. As a result, the measurement accuracy is less vulnerable to parameter drifts. Furthermore, the intrinsic redundancy of the setup can be used to check for systematic errors of measurement. The offset as well as scaling and directional errors of the magnetometer have been obtained by the method of scalar calibration in our laboratory, resulting in a residuum with a standard deviation of less than 0.64 nT. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1361-6501/28/1/015107

Additional details

Publishing Information

Journal Title
Measurement Science and Technology
Journal Volume
28
Journal Issue
1
Journal Page Range
[7 p.]
ISSN
0957-0233
CODEN