Description

The entrance skin dose (ESD) of chest X-ray examinations for AP and PA projections of paediatric patients at Maternity and Children Hospital in Najran, Saudi Arabia have been obtained using DoseCal software. The majority of the results obtained show low measured ESD for chest X-ray examinations. The mean of ESD for the AP projection was found to be 37.5, 40.5, 41.3, and 52.3μGy for age groups 0−1, >1−5, >5−10, and >10−15 years respectively. However, the ESD for PA projection was found to be 50.7 and 56.7μGy for age groups >5−10, and >10−15 years respectively.

Availability note (English)

Available from http://rd.springer.com/article/10.1007/s13246-012-0128-2

Additional details

Publishing Information

Journal Title
Australasian Physical and Engineering Sciences in Medicine
Journal Volume
35
Journal Issue
2
Journal Page Range
p. 215-219
ISSN
0158-9938

Optional Information

Notes
2 figa., 4 tabs.